Home
Pressure Calibration Kit (0-20KN): NIST certificated Loadcell + Digital Meter
|
Please email to sales@batteryspace.com for lead time
Product ID # 9144 Part Number: BS-070
|
BS-070 is designed to calibrate the crushing tester for UN38.3 and IEC62133. This kit includes a Compression Load Cell and a Digital Strain Gage Panel Meter. The compression load cell can be utilized between 0 to 20kN in compression force. Furthermore, the stainless steel construction ensures long term stability . The Miniature size and low profile allow the users to calibrate other load cell in a limited space or a small chamber.
The Digital Strain Gage Panel Meter is capable to be indicator or controller of the load cell with 4- digit standard LED display. The scaling capability of the meter also allows to display all engineering units. The wide DC current input ranges cover typical strain applications. The front panel includes several buttons to program, zero, and record the measurement.
Load Cell SPECIFICATIONS
Excitation: |
10 Vdc |
Output: |
2 mV/V nominal |
5-Point Calibration: |
0%, 50% 100%, 50%, 0% |
Linearity: |
±0.25% FSO |
Hysteresis: |
±0.20% FSO |
Repeatability: |
±0.10% FSO |
Zero Balance: |
±2% FSO |
Deflection: |
0.025 to 0.075 mm |
Operating Temp Range: |
-54 to 121°C (-65 to 250°F) |
Compensated Temp Range: |
16 to 71°C (60 to 160°F) |
Thermal Effects: |
Span: ±0.018% of Rdg/°C; Zero: ±0.009% of FSO/°C |
Safe Overload: |
150% of Capacity |
Ultimate Overload: |
300% of Capacity |
Bridge Resistance: |
350 Ohms minimum |
Construction: |
Stainless Steel |
Electrical Connection: |
1.5 m (5 ft.) Shielded Cable |
Digital Strain Gage Panel Meter Specification:
Input Ranges |
0-100 mV, ±50 mV, 0-10 V, ±5 V, 0-20 mA, 4-20 mA |
Isolation |
Dielectric strength to 2500V transient per 3mm spacing based on EN 61010 for 260Vrms or DC working voltage |
Input Protection |
Voltage Input = 120 Vrms max; Current Input = 100 mA max |
Input Impedance |
100 Meg ohms for 100 mV or ±50 mV input range; 1 Meg ohm for 10 V or ±5 V input range; 5 ohms for 20 mA current input range |
Display |
4-digit, LED 9-segment, 21 mm (0.83") |
Analog-to-digital technique |
Dual slope |
Internal resolution |
15 bits |
Read rate |
3/sec |
Polarity |
Automatic |
Max error process |
±0.03% of reading, ±1 count |
Span tempco |
50 ppm/°C |
Step response |
1 sec |
Warm-up to rated accuracy |
30 min |
Load Regulation |
1.1% |
Line Regulation |
0.02% per Vac |
Analog output (optional) |
0-10 V, 4-20 mA or 0-20 mA; may be assigned to a display range or to setpoint #1, when used as a proportional control output |
Operating temperature |
0 to 50°C (32 to 122°F) |
Storage temperature |
-40 to 85°C (-40 to 184°F) |
Relative humidity |
90% at 40°C (104°F) (non-condensing) |
Panel cutout |
1/8 DIN, 45 x 92 mm (1.78" x 3.62") |
Weight |
575 g (1.27 lb) |
AC Power Unit |
- Power: 115 Vac or 230 Vac ±10%; 50-60 Hz, 11.5 watts max.
- Excitation: 24 V @ 25 mA or 12 V @ 50 mA or 10 V @ 120 mA or 5 V @ 60 mA
|
DC Power Unit |
- Power: 10-32 Vdc; 7.5 watts max.
- Excitation: 24 V @ 25 mA or 12 V, 10 V, 5 V @ 35 mA (For Low/Medium Brightness)
|
Certification |
UL Certified |
|